Published November 1988
| Version v1
Journal article
Efficiency calibration and summation effects in gamma-ray spectrometry
Creators
- 1. Bilkent Univ., Ankara (Turkey). Dept. of Chemistry
- 2. Middle East Technical Univ., Ankara (Turkey)
Description
Precise, absolute γ-ray efficiencies (εγ) and total efficiencies (εT) were measured at various distances from the detector, using a set of standards. The observed linear variation of εγ and εT as well as εγ/εT with Eγ and 1/√d provides a simple means of interpolation and extrapolation. Experimental coincidence summation effects were determined for various nuclides and compared with calculated values. The results are found to be in good agreement with each other. (author) 12 refs.; 6 figs.; 2 tabs
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 125
- Journal Issue
- 1
- Series
- J. Radioanal. Nucl. Chem. ;Articles.
- Journal Page Range
- 3-10
- ISSN
- 0236-5731
- CODEN
- JRNCD
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 20053053
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CALIBRATION STANDARDS; COMPARATIVE EVALUATIONS; DISTANCE; EFFICIENCY; ENERGY DEPENDENCE; ENERGY LEVELS; GAMMA DETECTION; GAMMA SPECTROSCOPY; LI-DRIFTED GE DETECTORS; RADIOISOTOPES; VARIATIONS
- Descriptors DEC
- DETECTION; EVALUATION; GE SEMICONDUCTOR DETECTORS; ISOTOPES; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROSCOPY