Analytical Validation of a Portable Mass Spectrometer Featuring Interchangeable, Ambient Ionization Sources for High Throughput Forensic Evidence Screening
Creators
- 1. Illinois State University, Department of Chemistry (United States)
- 2. Illinois State University, Department of Management and Quantitative Methods (United States)
- 3. Innsbruck Medical University, Institute of Legal Medicine and Core Facility Metabolomics (Austria)
- 4. Illinois State University, Department of Criminal Justice Sciences (United States)
Description
Forensic evidentiary backlogs are indicative of the growing need for cost-effective, high-throughput instrumental methods. One such emerging technology that shows high promise in meeting this demand while also allowing on-site forensic investigation is portable mass spectrometric (MS) instrumentation, particularly that which enables the coupling to ambient ionization techniques. While the benefits of rapid, on-site screening of contraband can be anticipated, the inherent legal implications of field-collected data necessitates that the analytical performance of technology employed be commensurate with accepted techniques. To this end, comprehensive analytical validation studies are required before broad incorporation by forensic practitioners can be considered, and are the focus of this work. Pertinent performance characteristics such as throughput, selectivity, accuracy/precision, method robustness, and ruggedness have been investigated. Reliability in the form of false positive/negative response rates is also assessed, examining the effect of variables such as user training and experience level. To provide flexibility toward broad chemical evidence analysis, a suite of rapidly-interchangeable ion sources has been developed and characterized through the analysis of common illicit chemicals and emerging threats like substituted phenethylamines. .
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of the American Society for Mass Spectrometry
- Journal Volume
- 28
- Journal Issue
- 6
- Journal Page Range
- p. 1048-1059
- ISSN
- 1044-0305
- CODEN
- JAMSEF
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 50020286
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; FLEXIBILITY; ION SOURCES; IONIZATION; MASS SPECTROMETERS; MASS SPECTROSCOPY; PERFORMANCE; RELIABILITY; SCREENING
- Descriptors DEC
- MEASURING INSTRUMENTS; MECHANICAL PROPERTIES; SPECTROMETERS; SPECTROSCOPY; TENSILE PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2017 American Society for Mass Spectrometry