Bragg Curve Spectroscopy
Description
An alternative utilization is presented for the gaseous ionization chamber in the detection of energetic heavy ions, which is called Bragg Curve Spectroscopy (BCS). Conceptually, BCS involves using the maximum data available from the Bragg curve of the stopping heavy ion (HI) for purposes of identifying the particle and measuring its energy. A detector has been designed that measures the Bragg curve with high precision. From the Bragg curve the range from the length of the track, the total energy from the integral of the specific ionization over the track, the dE/dx from the specific ionization at the beginning of the track, and the Bragg peak from the maximum of the specific ionization of the HI are determined. This last signal measures the atomic number, Z, of the HI unambiguously
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS., PC A03/MF A01.
Files
Additional details
Publishing Information
- Imprint Pagination
- 27 p.
- Report number
- LBL--12678
Conference
- Title
- INS meeting on nuclear radiation detectors.
- Dates
- 23 - 26 Mar 1981.
- Place
- Tokyo, Japan.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 12639128
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BRAGG CURVE; HEAVY ION SPECTROMETERS; PERFORMANCE; RESOLUTION
- Descriptors DEC
- DIAGRAMS; INFORMATION; MEASURING INSTRUMENTS; SPECTROMETERS
Optional Information
- Secondary number(s)
- CONF-810351--3.