Bragg Curve Spectroscopy

Description

An alternative utilization is presented for the gaseous ionization chamber in the detection of energetic heavy ions, which is called Bragg Curve Spectroscopy (BCS). Conceptually, BCS involves using the maximum data available from the Bragg curve of the stopping heavy ion (HI) for purposes of identifying the particle and measuring its energy. A detector has been designed that measures the Bragg curve with high precision. From the Bragg curve the range from the length of the track, the total energy from the integral of the specific ionization over the track, the dE/dx from the specific ionization at the beginning of the track, and the Bragg peak from the maximum of the specific ionization of the HI are determined. This last signal measures the atomic number, Z, of the HI unambiguously

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A03/MF A01.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
27 p.
Report number
LBL--12678

Conference

Title
INS meeting on nuclear radiation detectors.
Dates
23 - 26 Mar 1981.
Place
Tokyo, Japan.

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12639128
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BRAGG CURVE; HEAVY ION SPECTROMETERS; PERFORMANCE; RESOLUTION
Descriptors DEC
DIAGRAMS; INFORMATION; MEASURING INSTRUMENTS; SPECTROMETERS

Optional Information

Secondary number(s)
CONF-810351--3.