Published March 1, 2021 | Version v1
Journal article

Fading estimation of SOL-GEL α-Al2O3 detectors

  • 1. Instituto de Pesquisas Energéticas e Nucleares/Comissão Nacional de Energia Nuclear, IPEN/CNEN, Av. Prof. Lineu Prestes, 2242, 05508-000, São Paulo, SP (Brazil)

Description

The OSL and the TL techniques are used in several fields of radiation dosimetry. The dosimeters are affected by a signal loss in the time between the irradiation and readout (fading). This undesirable characteristic can cause an underestimation of the irradiation dose. The fading estimation of SOL-GEL α-Al2O3 with several different concentrations of impurities is reported. The thermal fading and the light-induced fading of the SOL-GEL α-Al2O3 detectors were estimated. Furthermore, the exponential equations used to interpolate the experimental results were presented. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/1826/1/012045

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
1826
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53079310
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ALUMINIUM OXIDES; DOSEMETERS; DOSIMETRY; IRRADIATION; RADIATION DOSES; READOUT SYSTEMS; SIGNALS; SOL-GEL PROCESS
Descriptors DEC
ALUMINIUM COMPOUNDS; CHALCOGENIDES; DOSES; MEASURING INSTRUMENTS; OXIDES; OXYGEN COMPOUNDS