Published 2004 | Version v1
Journal article

A tool-kit for epithermal neutron beam characterisation in BNCT

  • 1. VTT Processes, VTT Technical Res. Centre of Finland, P.O. Box 1608, FIN-02044 VTT (Finland)
  • 2. Dept. of Physical Sciences, Univ. of Helsinki, P.O. Box 64, FIN-00014 Helsinki (Finland)
  • 3. Helsinki Medical Imaging Center, Univ. of Helsinki, P.O. Box 340, FIN-00029 HUCS (Finland)
  • 4. Radiation Metrology Laboratory, Radiation/Nuclear Safety Authority, P.O. Box 14, FIN-00881 Helsinki (Finland)

Description

Methods for dosimetry of epithermal neutron beams used in boron neutron capture therapy (BNCT) have been developed and utilised within the Finnish BNCT project as well as within a European project for a code of practise for the dosimetry of BNCT. One outcome has been a travelling tool-kit for BNCT dosimetry. It consists of activation detectors and ionisation chambers. The free-beam neutron spectrum is measured with a set of activation foils of different isotopes irradiated both in a Cd-capsule and without it. Neutron flux (thermal and epithermal) distribution in phantoms is measured using activation of Mn and Au foils, and Cu wire. Ionisation chamber (IC) measurements are performed both in-free-beam and in-phantom for determination of the neutron and gamma dose components. This tool-kit has also been used at other BNCT facilities in Europe, the USA, Argentina and Japan. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch151

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
110
Journal Issue
1-4
Journal Page Range
p. 587-593
ISSN
0144-8420

Conference

Title
9. Symposium on Neutron Dosimetry - Advances in nuclear particle dosimetry for radiation protection and medicine
Dates
28 Sep - 3 Oct 2003
Place
Delft (Netherlands)

Optional Information

Notes
24 refs.