Portable detectors for 125I-insulin absorption measurement during subcutaneous infusion with portable pumps
Creators
- 1. Finseninstitutet, Copenhagen (Denmark)
- 2. Steno Memorial Hospital, Copenhagen (Denmark)
Description
Programmed subcutaneous insulin infusion is a promising method for normalisation of the blood glucose concentration in insulin-dependent diabetics. The absorption rate from the depot is usually measured intermittently by radioactively-labelled insulin and stationary scintillation detectors. Small portable detectors are an alternative, however, and continuous absorption measurements could be made during normal life conditions. Contrary to conventional single injection therapy, the insulin depot initially expands during infusion treatment, changing the geometry during measurements. In the present study the methodological aspects and geometrical dependences were investigated. Simulated studies were made with various plane disc 125I sources in Perspex phantoms as well as 125I-insulin absorption studies in short-term subcutaneous infusion experiments with anaesthetised rabbits. Results from portable, end-window Geiger-Mueller (GM) detectors fixed above the depots and close to the surfaces of phantom or skin were compared with results obtained by a conventional stationary NaI(Tl) detector 15 cm from the phantom or skin surface. With a 125I-insulin infusion site at 5 mm depth in the subcutaneous tissue of rabbits, an overall linear proportionality was found between the results obtained with a NaI(Tl) detector and a GM detector raised 15 mm above the skin surface inside the detector housing. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 29
- Journal Issue
- 4
- Series
- Phys. Med. Biol.
- Journal Page Range
- 395-405
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 15056933
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DIABETES MELLITUS; GEIGER-MUELLER COUNTERS; INSULIN; IODINE 125; NAI DETECTORS; PHANTOMS; SUBCUTANEOUS INJECTION; UPTAKE
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; DISEASES; ELECTRON CAPTURE RADIOISOTOPES; ENDOCRINE DISEASES; HORMONES; INJECTION; INTAKE; INTERMEDIATE MASS NUCLEI; IODINE ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MOCKUP; NUCLEI; ODD-EVEN NUCLEI; PEPTIDE HORMONES; RADIATION DETECTORS; RADIOISOTOPES; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; STRUCTURAL MODELS