Published February 15, 2006 | Version v1
Miscellaneous

Volume Optimization for Activity Determination of Tritiated Water Samples by Liquid Scintillation Counting

  • 1. Radiation Protection, Nuclear Research Center Negev, Beer Sheva (Israel)

Description

The method for determination of tritium activity in water samples by Liquid Scintillation Counting has been validated in our laboratory for accreditation to ISO 17025. The validation process included a renewed examination of the method with relation to the accuracy, precision, selectivity, stability, linearity and detection limit. The present work focused on determination of the optimal volumes for the samples preparation. It included the optimization of the total volume of the sample, and of the ratio between the volume of the sample and of the liquid scintillator. For determination of the optimal values, we have tested several parameters: the counting efficiency (which depends on the quenching in the sample and the counting geometry), the background and the minimum detectable activity

Part of:
The 23. conference of the Nuclear Societies in Israel. Book of articles

Additional details

Publishing Information

Imprint Title
The 23. conference of the Nuclear Societies in Israel. Book of articles
Imprint Pagination
256 p.
Journal Page Range
p. 72-74
Report number
INIS-IL--012

Conference

Title
23. conference of the Nuclear Societies in Israel
Dates
15-16 Feb 2006
Place
Dead Sea (Israel)

Optional Information