Electrocardiographic gating in positron emission computed tomography

  • 1. Univ. of California, Los Angeles

Description

Electrocardiographic (ECG) synchronized multiple gated data acquisition was employed with positron emission computed tomography (ECT) to obtain images of myocardial blood pool and myocardium. The feasibility and requirements of multiple gated data acquisition in positron ECT were investigated for 13NH3, (18F)-2-fluoro-2-D-deoxyglucose, and (11C)-carboxyhemoglobin. Examples are shown in which image detail is enhanced and image interpretation is facilitated when ECG gating is employed in the data collection. Analysis of count rate data from a series of volunteers indicates that multiple, statistically adequate images can be obtained under a multiple gated data collection format without an increase in administered dose

Additional details

Publishing Information

Journal Title
J. Comput. Assist. Tomogr.
Journal Volume
3
Journal Issue
6
Series
J. Comput. Assist. Tomogr.
Journal Page Range
733-739