Disposable blood potassium sensors based on screen-printed thick film electrodes

  • 1. State Key Laboratory of Industrial Control Technology, Institute of Cyber Systems and Control, Zhejiang University, Hangzhou 310027 (China)
  • 2. Department of Biomedical Engineering, Zhejiang University, Hangzhou 310027 (China)
  • 3. College of Informatics and Electronics, Zhejiang Sci-Tech University, Hangzhou, 310018 (China)
  • 4. Computer Learning Research Centre, Royal Holloway, University of London, Egham, Surrey TW20 0EX (United Kingdom)

Description

A sensor based on the screen-printing technique for the rapid blood potassium test was developed. The sensor consists of a thick film potassium ion-selective electrode and a reference electrode. Poly(3,4-ethylenedioxythiophene) (PEDOT) doped with poly(styrenesulfonate) (PSS) was screen-printed on carbon electrodes as an internal solid contact layer for both ion-selective and reference electrodes. PVC membranes with and without selective ionphore valinomycin were coated on the PEDOT/PSS layer to form potassium ion-selective and reference electrodes, respectively. The potentiometric response characters were evaluated. The response time of the sensor was 15 s when less than 15 µL of sample was applied. The sensitivity of the potassium sensor was around 60 mV per decade, while the potentiometric log selectivity coefficients were −3.43 and −3.48 for Na+ and Ca2+, respectively

Availability note (English)

Available from http://dx.doi.org/10.1088/0957-0233/21/5/055802

Additional details

Identifiers

DOI
10.1088/0957-0233/21/5/055802;
PII
S0957-0233(10)40093-4;

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
21
Journal Issue
5
Journal Page Range
[5 p.]
ISSN
0957-0233
CODEN
MSTCEP