Fricke gel dosimeter as a tool in quality assurance of the radiotherapy treatment plans
- 1. National Center for Radiation Research and Technology, Atomic Energy Authority (AEA), P.O. Box 8029, Nasr City, Cairo (Egypt)
- 2. Biophysics Branch, Physics Department, Faculty of Science, Al-Azhar University, Nasr City, Cairo (Egypt)
- 3. Medical physics unit, Ain-Shams hospital, Ain-Shams University, ElAbbasia, Cairo (Egypt)
Description
Ferrous xylenol gel was evaluated as a dosimeter in verifications of treatment plans for treatments of female breast tumors in external-beam radiation therapy. The dosimeter was calibrated in a cubic wax phantom irradiated with 6 and 15 MV beams of an Elekta clinical linear accelerator and then used for measurements of doses in an anthropomorphic phantom of a female torso, which mimicked female breasts after modified radical mastectomy and breast-conserving surgery. Doses measured with the gel in specific locations in the phantom were compared with values predicted by treatment plans created with the XiO treatment planning system (Elekta / IMPAQ) and with results of measurements with an IBA CC13 ionization chamber in the same locations. Differences between the values measured with the gel and predicted by the treatment planning system or measured with the ionization chamber were within 1.5% in most cases. In an area of as steep dose gradient, the difference reached 2.7%, which was explicable in terms of the finite size of the gel dosimeter. The dose response of the gel is dose rate and energy independent in the ranges used in most clinical linacs. The results have shown that ferrous xylenol gels can be used as dosimeters in quality assurance in radiation therapy. - Highlights: • Fricke gel dosimetry is suitable for verifications of radiotherapy treatment plans. • Differences between its results and ionization chamber readings or doses predicted by a TPS are within 1.5% in most cases. • The dose response is energy and dose-rate independent.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.apradiso.2016.12.004Additional details
Identifiers
- DOI
- 10.1016/j.apradiso.2016.12.004;
- PII
- S0969-8043(16)31056-9;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 120
- Journal Page Range
- p. 126-132
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48089600
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DOSE RATES; DOSEMETERS; DOSIMETRY; EXTERNAL BEAM RADIATION THERAPY; FEMALES; GELS; IONIZATION; IONIZATION CHAMBERS; IRRADIATION; LINEAR ACCELERATORS; MAMMARY GLANDS; NEOPLASMS; PHANTOMS; QUALITY ASSURANCE; RADIATION DOSES; RADICALS; SURGERY; XYLENOLS
- Descriptors DEC
- ACCELERATORS; AROMATICS; BODY; COLLOIDS; DISEASES; DISPERSIONS; DOSES; EVALUATION; GLANDS; HYDROCARBONS; HYDROXY COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; ORGANS; PHENOLS; RADIATION DETECTORS; RADIOLOGY; RADIOTHERAPY; STRUCTURAL MODELS; THERAPY
Optional Information
- Copyright
- Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.