Published February 1997 | Version v1
Journal article

Advantage of fan beam collimators for contrast recovery of hyperfixation in clinical SPECT

  • 1. Hopital Charles Nicolle, Rouen (France). Dept. of Biophysics
  • 2. Service Hospitalier Frederic Joliot, Orsay (France)
  • 3. Hopital Beaujon, Clichy (France). Dept. of Nuclear Medicine

Description

The influence of the collimator on the contrast recovery of hyperfixation was studied using a dual-headed single photon emission computer tomography (SPECT) system with standard clinical acquisition parameters. Three parallel collimator sets and two fan beam collimator sets were tested with a Jaszczak phantom. The six spheres of the phantom were filled with 99mTc, and four background levels were progressively obtained by adding radioactivity to the cylinder of the phantom, providing four hyperfixation levels. The effects of angular sampling and reconstruction filters have been tested. The statistical analysis was performed with analysis of variance (ANOVA). This study demonstrates the advantage of ultra-high resolution fan beam collimators for contrast recovery of hyperfixation with SPECT when using 64 projections over 360 degree, in particular when the contrast is low. The authors also demonstrate that fan beam collimators permit smaller size hyperfixation detection

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
44
Journal Issue
1
Journal Page Range
p. 83-89.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
IEEE nuclear science symposium and medical imaging conference.
Dates
23-28 Oct 1995.
Place
San Francisco, CA (United States).

Optional Information

Secondary number(s)
CONF-951073--.