Published December 1990
| Version v1
Report
Neutron scattering from N-oxy γ-picoline
Description
Neutron scattering experiments on N-oxy γ-picoline have been performed by using high energy-resolution spectrometer, LAM-80ET. An aim of this experiment was to check the performance of LAM-80ET and a feasibility on the observation of the rotational tunneling. By measurements on vanadium and N-oxy γ-picoline samples, it was confirmed that LAM-80ET has extremely high energy-resolution; ∼1 μeV by (002) reflection of the mica crystal analyzer; ∼5 μeV by (004); ∼15 μeV by (006). Very clear tunnel mode's peaks of N-oxy γ-picoline were observed at 60, 115, 190 and 274 μeV at 5 K. (author)
Additional details
Publishing Information
- Imprint Title
- TUNNEL 902
- Imprint Pagination
- 149 p.
- Journal Page Range
- p. 42-58.
- Report number
- KEK--90-17
Conference
- Title
- 2. meeting on tunnel.
- Dates
- 24 Apr 1990.
- Place
- Tsukuba, Ibaraki (Japan).
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 22069751
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COLD NEUTRONS; ELASTIC SCATTERING; ENERGY RESOLUTION; INELASTIC SCATTERING; KEK SYNCHROTRON; NEUTRON SOURCES; NEUTRON SPECTROMETERS; TEMPERATURE DEPENDENCE; TIME-OF-FLIGHT METHOD; TUNNEL EFFECT
- Descriptors DEC
- ACCELERATORS; BARYONS; CYCLIC ACCELERATORS; ELEMENTARY PARTICLES; FERMIONS; HADRONS; MEASURING INSTRUMENTS; NEUTRONS; NUCLEONS; PARTICLE SOURCES; RADIATION SOURCES; RESOLUTION; SCATTERING; SPECTROMETERS; SYNCHROTRONS