Pile-up effect for a counting channel with amplitude selection and imposed non-cumulative dead time
Creators
- 1. Institut de Radiophys. Appliquee et Office Federal de Metrologie (IRA-OFMET), Lausanne (Switzerland)
Description
The performance of radioactivity counting systems is limited by pile-up which causes counting losses, timing uncertainties and incorrect energy selection. The counting signals are either single pulses for which amplitude distribution is observable at low count rate, or more complex signals for which the amplitude distribution is, for the superposition of rectangular pulses, the self convolution of the single-pulse spectrum. Without amplitude selection, the effect on the count rate is small. It is the same as an extending dead time equal to the width of a single pulse. However, amplitude selection reveals a more important distortion. The output count rate has been derived for a single channel analyser followed by a non-cumulative dead time. These results are confirmed by computer simulations. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 369
- Journal Issue
- 2-3
- Journal Page Range
- p. 397-400.
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- International symposium on radionuclide metrology and its applications.
- Original Conference Title
- Symposium International sur la Metrologie des Radionucleides et ses Applications
- Dates
- 15-19 May 1995.
- Place
- Paris (France).
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 27032101
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMPLITUDES; COMPUTERIZED SIMULATION; COUNTING CIRCUITS; COUNTING RATES; DEAD TIME; DISTURBANCES; ELECTRON CAPTURE DECAY; ENERGY SPECTRA; GAMMA SPECTRA; GAMMA SPECTROSCOPY; MANGANESE 54; PULSE ANALYZERS; PULSE DISCRIMINATORS; PULSE PILEUP
- Descriptors DEC
- BETA DECAY; BETA DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; DECAY; DISCRIMINATORS; ELECTRON CAPTURE RADIOISOTOPES; ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; INTERMEDIATE MASS NUCLEI; ISOTOPES; MANGANESE ISOTOPES; NUCLEAR DECAY; NUCLEI; ODD-ODD NUCLEI; PULSE CIRCUITS; RADIOISOTOPES; SIMULATION; SPECTRA; SPECTROSCOPY; TIMING PROPERTIES