Published August 2014 | Version v1
Miscellaneous Open

Tl dosimetry around equipment s for dental radiography

  • 1. Universidad Autonoma de Zacatecas, Unidad Academica de Odontologia, Apdo. Postal 336, 98000 Zacatecas (Mexico)
  • 2. Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico)

Description

The ambient dose equivalent around X-rays units used in dentistry has been measured. The radiation that escapes of the tail stock of the X-rays equipment was also measured, and the dose in the surfaces of a polyethylene spherical mannequin and other of water, used to represent the patient and the operator, respectively. The study was realized with thermoluminescent dosimeters of the type 100. The operation conditions of the X-rays units were varied when the ambient dose equivalent was measured. In the other measurements the equipment s were operated to 70 k Vp and 8 m A. In the ambient dose equivalent was found asymmetry, likewise the due dose to the radiation that escapes of one of tail stocks was also asymmetric. The dose values in the mannequins went minors to the recommended reference values. (Author)

Files

45089762.pdf

Files (458.9 kB)

Name Size Download all
md5:6c715a5fef1a564b22aec2f19f2c7ad6
458.9 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Dosimetria TLD en torno a equipos para radiografia dental

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico, D. F. (Mexico)
Imprint Pagination
17 p.
Report number
INIS-MX--2729

Conference

Title
14. International Symposium on Solid State Dosimetry
Original Conference Title
14. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
13-16 Apr 2014
Place