Published May 2016 | Version v1
Journal article

A Monte Carlo-based dosimetric study of the GZP 60Co source

  • 1. Department of Radiotherapy, Sichuan Cancer Hospital, Chengdu (China)

Description

Objective: To simulate and calculate the dosimetric parameters of the GZP 60Co source that has been clinically used in high-dose-rate brachytherapy. Methods: The EGSnrc Monte Carlo software was used to simulate and calculate the dosimetric parameters of a well known BEBIG 60Co source (Co0.A86). The results were compared with the actual parameters to verify the feasibility of this method. A Monte Carlo model of the GZP 60Co source for high-dose-rate brachytherapy was established to simulate and calculate its dosimetric parameters in the same way. Results: For the BEBIG 60Co source, the results were well accorded with the standard. The air-kerma strength per unit activity (SK/A) and dose rate constant (A) deviated from the standard by 0.2% and 1.0%, respectively. The curves of the radial dose function gL(r) and the anisotropy function F(r, θ) fit well. For the GZP 60Co source, the SK/A and Λ values were calculated as 3.011 × 10-7 cGycm2h-1Bq-1 and 1.118 cGyh-1U-1 in channel 1&2 and 3.002 × 10-7 cGycm2h-1Bq-1 and 1.110 cGyh-1U-1 in channel 3. The gL(r), F (r, 0θ), and dose rate per unit air-kerma strength in a liquid water phantom were listed as the American Association of Physicists in Medicine recommended. Conclusions: The results can be used in planning system and quality control for the GZP 60Co source. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiation Oncology
Journal Volume
25
Journal Issue
5
Journal Page Range
p. 489-495
ISSN
1004-4221

Optional Information

Notes
4 figs., 7 tabs., 11 refs.; http://dx.doi.org/10.3760/cma.j.issn.1004-4221.2016.05.015