Preparation of glucose sensors using gold nanoparticles modified diamond electrode

  • 1. Department of Chemistry, Faculty of Mathematics and Natural Sciences Universitas Indonesia, Kampus UI Depok, Depok 16424 (Indonesia)

Description

A glucose sensor was successfully developed by immobilizing glucose oxidase (GOx) at boron-doped diamond (BDD) electrodes. Prior to GOx immobilization, the BDD was modified with gold nanoparticles (AuNPs). To immobilize AuNPs, the gold surface was modified to nitrogen termination. The characterization of the electrode surface was performed using an X-ray photoelectron spectroscopy and a scanning electron microscope, while the electrochemical properties of the enzyme electrode were characterized using cyclic voltammetry. Cyclic voltammograms of the prepared electrode for D-glucose in phosphate buffer solution pH 7 showed a new reduction peak at +0.16 V. The currents of the peak were linear in the concentration range of 0.1 M to 0.9 M, indicated that the GOx-AuNP-BDD can be applied for electrochemical glucose detection. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1757-899X/188/1/012046

Additional details

Publishing Information

Journal Title
IOP Conference Series. Materials Science and Engineering (Online)
Journal Volume
188
Journal Issue
1
Journal Page Range
[5 p.]
ISSN
1757-899X

Conference

Title
International symposium on current progress in functional materials
Dates
26-27 Jul 2016
Place
Bali (Indonesia)