Determination of activity meter settings for the PET nuclides 44Sc and 89Zr

  • 1. Laboratorio de Metrología de Radiaciones Ionizantes, CIEMAT (Spain)
  • 2. Unidad de Aplicaciones Biomédicas y Farmacocinética, CIEMAT Avda, Complutense 40, Madrid, 28040 (Spain)

Description

Highlights: • Primary standardisation of 44Sc and 89Zr performed. • Reported dial settings for radionuclide calibrators in clinically relevant geometries. • New dial settings identified for calibrators and clinically relevant geometries. • Results provide data that did not previously exist in the literature. - Abstract: Radionuclide activity meters depend on "calibration setting numbers" that are specific for a certain combination of radionuclide, volume and geometry. This work describes their determination for two PET emerging nuclides (44Sc and 89Zr), three commercial activity meters (CAPINTEC CRC15 BETA, ATOMLAB BIODEX 500 and VEENSTRA VDC 405) and two geometries: 1 mL plastic syringes and 5 mL glass vials. For 44Sc, only values for the dial setting for 5 mL ampoules were reported previously and its use for the assayed geometries would lead to significant differences. For 89Zr more data were available in the literature, and results are provided for new geometries. All results rely on the absolute standardisation in activity of solutions of both radionuclides.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2019.108829

Additional details

Identifiers

DOI
10.1016/j.apradiso.2019.108829;
PII
S0969804319301083;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
153
Journal Page Range
p. 108829
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Notes
© 2019 Elsevier Ltd. All rights reserved.