Published April 2017 | Version v1
Journal article

Ultra-high resolution C-Arm CT arthrography of the wrist: Radiation dose and image quality compared to conventional multidetector computed tomography

  • 1. Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover (Germany)
  • 2. University of Veterinary Medicine Hannover, Institute for General Radiology and Medical Physics, Bischofsholer Damm 15, 30173 Hannover (Germany)

Description

Objective: Objective of this phantom and cadaveric study was to compare the effective radiation dose (ED) and image quality (IQ) between C-arm computed tomography (CACT) using an ultra-high resolution 1 × 1 binning with a standard 16-slice CT (MDCT) arthrography of the wrist. Methods: ED was determined with thermoluminescence dosimetry using an anthropomorphic phantom and different patient positions. Imaging was conducted in 10 human cadaveric wrists after tri-compartmental injection of diluted iodinated contrast material and a wire phantom. IQ of MDCT was compared with CACT reconstructed with a soft (CACT1) and sharp (CACT2) kernel. High and low contrast resolution was determined. Three radiologists assessed IQ of wrist structures and occurrence of image artifacts using a 5-point Likert scale. Results: ED of MDCT was comparable to standard CACT (4.3 μSv/3.7 μSv). High contrast resolution was best for CACT2, decreased to CACT1 and MDCT. Low contrast resolution increased between CACT2 and MDCT (P < 0.001). IQ was best for CACT2 (1.3 ± 0.5), decreased to CACT1 (1.9 ± 0.6) and MDCT (3.5 ± 0.6). Non-compromising artifacts were only reported for CACT. Conclusions: The results of this phantom and cadaveric study indicate that ultra-high resolution C-Arm CT arthrography of the wrist bears the potential to outperform MDCT arthrography in terms of image quality and workflow at the cost of mildly increasing image artifacts while radiation dose to the patient is comparably low for both, MDCT and C-Arm CT.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2017.02.009

Additional details

Identifiers

DOI
10.1016/j.ejrad.2017.02.009;
PII
S0720-048X(17)30052-9;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
89
Journal Page Range
p. 191-199
ISSN
0720-048X
CODEN
EJRADR

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49011443
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; IMAGES; PHANTOMS; RADIATION DOSES; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
DIAGNOSTIC TECHNIQUES; DOSES; DOSIMETRY; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; TOMOGRAPHY