Published 1976 | Version v1
Book

A TLD method for the determination of x-ray quality and output in the diagnostic range

  • 1. Bhabha Atomic Research Centre, Bombay (India). Div. of Radiation Protection

Description

A simple TLD method is described for the simultaneous determination of NVT and output of diagnostic X-ray machines. Thermoluminescent Li2B4O7: Mn phosphor sealed in polythene bags and attached to four 0.7 mm thick aluminium sheets as a step wedge was used as a compact packet and exposed to various energies of X-rays. The TL output was measured and plotted against thickness of Al and the HVT determined. Comparisons made with the usual aluminium wedge and X-ray dosimeter data showed agreement with +- 5% in the range of 40 to 130 kV. The HVT-TL- dosimeter wedge is convenient for handling and suitable especially for making measurements remotely by mail. (author)

Additional details

Publishing Information

Publisher
Bhabha Atomic Research Centre.
Imprint Place
Bombay
Imprint Title
Proceedings of the national symposium on thermoluminescence and its applications, Kalpakkam, February 12-15, 1975
Imprint Pagination
p. 469-474.

Conference

Title
National symposium on thermoluminescence and its applications.
Dates
12 Feb 1975.
Place
Kalpakkam, India.

Optional Information

Notes
7 refs.