Published November 2018 | Version v1
Journal article

Comparision of pre-treatment EPID based patient specific QA results with during-treatment trajectory log files analysis of truebeam linear accelerator

  • 1. Department of Radiation Oncology, National Cancer Institute, Nagpur (India)

Description

IMRT (Intensity Modulated Radiotherapy)/VMAT (Volumetric Modulated Radiotherapy) is a complex process which requires a proper QA programme in place in order to ensure a accurate tumoricidal dose delivery. The accuracy of dose delivery must be documented for each course of treatment by irradiating a phantom that contains array of detectors to sample the dose distribution and to verify that the dose delivered is the dose planned

Additional details

Publishing Information

Journal Title