Relative Patlak plot for dynamic PET parametric imaging without the need for early-time input function
Creators
- 1. Department of Radiology, University of California at Davis, Sacramento, CA 95817 (United States)
- 2. Department of Biomedical Engineering, University of California at Davis, Davis, CA 95616 (United States)
Description
The Patlak graphical method is widely used in parametric imaging for modeling irreversible radiotracer kinetics in dynamic PET. The net influx rate of radiotracer can be determined from the slope of the Patlak plot. The implementation of the standard Patlak method requires the knowledge of full-time input function from the injection time until the scan end time, which presents a challenge for use in the clinic. This paper proposes a new relative Patlak plot method that does not require early-time input function and therefore can be more efficient for parametric imaging. Theoretical analysis proves that the effect of early-time input function is a constant scaling factor on the Patlak slope estimation. Thus, the parametric image of the slope of the relative Patlak plot is related to the parametric image of standard Patlak slope by a global scaling factor. This theoretical finding has been further demonstrated by computer simulation and real patient data. The study indicates that parametric imaging of the relative Patlak slope can be used as a substitute of parametric imaging of standard Patlak slope for tasks that do not require absolute quantification, such as lesion detection and tumor volume segmentation. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/aad444Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 63
- Journal Issue
- 16
- Journal Page Range
- [12 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52003073
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- INJECTION; KINETICS; NEOPLASMS; PATIENTS; POSITRON COMPUTED TOMOGRAPHY; TRACER TECHNIQUES
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; EMISSION COMPUTED TOMOGRAPHY; INTAKE; ISOTOPE APPLICATIONS; TOMOGRAPHY