Published 1991
| Version v1
Journal article
Automatic control of uranium and nitric acid in reextraction solutions
Description
New indirect potentiometric method on the basis of two ion-selective electrodes EM-NO3-01 and ESL-43-07 is proposed for semultaneous determination of uranium and nitric acid under dynamic conditions. Application of the method with the simpliest equipment makes it possible to determine uranium and nitric acid with relative average quadratic deviation << 0.03. Area of uranium concentrations determined ranges from 59 to 117 g/l and of nitric acid - 2.5 to 4.1 g/l. Transducer for continuous automatic control of uranium and nitric acid in reextraction solutions is created on the basis of the studies performed
Additional details
Additional titles
- Original title (Russian)
- Автоматический контрол' урана и азотной кислоты в реэкстракционных растворах
Publishing Information
- Journal Title
- Radiokhimiya
- Journal Volume
- 33
- Journal Issue
- 5
- Journal Page Range
- p. 171-179.
- ISSN
- 0033-8311
- CODEN
- RADKAU
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- Russian Federation
- INIS RN
- 24000303
- Subject category
- S11: NUCLEAR FUEL CYCLE AND FUEL MATERIALS;
- Descriptors DEI
- ACCURACY; AQUEOUS SOLUTIONS; CONTROL EQUIPMENT; ION SELECTIVE ELECTRODE ANALYS; NITRIC ACID; POTENTIOMETRY; QUANTITATIVE CHEMICAL ANALYSIS; QUANTITY RATIO; SENSITIVITY; URANIUM NITRATES
- Descriptors DEC
- ACTINIDE COMPOUNDS; CHEMICAL ANALYSIS; DISPERSIONS; EQUIPMENT; HOMOGENEOUS MIXTURES; HYDROGEN COMPOUNDS; INORGANIC ACIDS; INORGANIC COMPOUNDS; MIXTURES; NITRATES; NITROGEN COMPOUNDS; OXYGEN COMPOUNDS; SOLUTIONS; TITRATION; URANIUM COMPOUNDS