Published November 15, 1990 | Version v1
Journal article

An extended range neutron rem counter

  • 1. Milan Univ., Segrate (Italy). Dipt. di Fisica
  • 2. Istituto Nazionale di Fisica Nucleare, Segrate (Italy)
  • 3. Istituto Nazionale di Fisica Nucleare, Frascati (Italy). Lab. Nazionale di Frascati
  • 4. Consiglio Nazionale delle Ricerche, Milan (Italy). Ist. Tecnologie Biomediche Avanzate

Description

Extensive Monte Carlo calculations have been carried out to assess the possibility of extending the sensitivity of a neutron rem counter of the Andersson-Braun type up to several hundred MeV. The validity of the model adopted has first been checked by comparing with experimental data the calculated response curve and the angular dependence of the sensitivity for a well known commercial rem counter. Next, a number of modifications to the configuration of the moderator-attenuator have been investigated. The response functions and angular distributions produced by two simple solutions yielding an instrument with a sensitivity extended up to 400 MeV are presented. The response of the original rem counter and of its two modified versions to nine test spectra has also been calculated. The resulting instrument is transportable rather than portable, but the availability of an extended range neutron survey meter would be of great advantage at medium and high energy particle accelerator facilities. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
297
Journal Issue
1/2
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
250-257
ISSN
0168-9002
CODEN