Published 2008 | Version v1
Journal article

Commissioning and constancy protocols for digital angiographic units

  • 1. Medical Physics Dept., Konstantopoulio Agia Olga Hospital, Athens (Greece)
  • 2. Servizio di Fisica Medica, Ospedale 'Santa Maria della Misericordia', Udine (Italy)
  • 3. Radiology Dept., Ciudad Universitaria, Complutense Univ., Madrid (Spain)
  • 4. Ministry of Health, Div. of Radiation Protection, Luxembourg (Luxembourg)
  • 5. Medical Physics Dept., Evangelismos Hospital, Athens (Greece)
  • 6. Medical Physics Dept., Athens General Hospital 'G. Gennimatas', Athens (Greece)
  • 7. Medical Physics Dept., Medical School, Univ. of Athens, Athens (Greece)

Description

During the European Concerted Action SENTINEL 'Safety and Efficacy for New Techniques and Imaging using New Equipment to Support European Legislation', protocols for commissioning and constancy tests for dynamic digital flat detectors angiography units have been developed in order to harmonise practice among the European counties. The commissioning protocol includes measurements on X-ray tube and generator, patient and detector radiation dose and image quality. The constancy protocol is based on the dose and image quality measurements. The commissioning protocol was tested by SENTINEL partners who expressed an interest in checking their dynamic digital systems using this protocol. The results of basic tests are reported. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncn016

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
129
Journal Issue
1-3
Journal Page Range
p. 258-260
ISSN
0144-8420

Conference

Title
EC SENTINEL Project workshop. Safety and efficacy for new techniques and imaging using new equipment to support European legislation
Dates
18-20 Apr 2007
Place
Delf (Netherlands)

Optional Information

Notes
5 refs