Published November 1975 | Version v1
Report Open

Frequency measurement of the g-Factor of tetracene+

Description

A high precision in g-factor measurements is achieved by placing the ESR sample in the centre of a NMR-probe containing glycerol doped with 0,2 m CuSO4. Both absorption lines are measured and stored simultaneously by means of a signal averager. The relation between field values and memeory addresses is obtained by slight shifting of the NMR oscillator frequency in each run. Applying this method to the measurement of the g-factor of tetracene+ yields g=2,0026355+-3,1 ppm referred to the free proton

Availability note (English)

MF available from INIS under the Report Number.

Files

8313844.pdf

Files (207.7 kB)

Name Size Download all
md5:a1aae128386f08e23b6cdc5e5bd4af1d
207.7 kB Preview Download

Additional details

Additional titles

Original title (German)
Praezisionsmessung des g-Faktors von Tetrazen

Publishing Information

Imprint Pagination
7 p.
Report number
AIAU--75110

INIS

Country of Publication
Austria
Country of Input or Organization
Austria
INIS RN
8313844
Subject category
S74: ATOMIC AND MOLECULAR PHYSICS;
Descriptors DEI
AROMATICS; ELECTRON SPIN RESONANCE; LANDE FACTOR; TETRACENE
Descriptors DEC
CONDENSED AROMATICS; HYDROCARBONS; MAGNETIC RESONANCE; ORGANIC COMPOUNDS; RESONANCE

Optional Information

Notes
11 refs.