Published March 7, 2004 | Version v1
Journal article

Irradiating cell samples in an x-ray radiation cabinet-the effect of tube filtration

  • 1. East Anglian Regional Radiation Protection Service, Addenbrooke's Hospital, Hills Road, Cambridge CB2 2QQ (United Kingdom)
  • 2. Hutchison Laboratory, Medical Research Council, Cambridge CB2 2XZ (United Kingdom)

Description

The cell irradiation cabinet described is used for creating DNA damage in cell samples in order to study tumourigenesis. The medical research laboratory involved was using the manufacturer's quoted dose rate (32.2 mGy s-1) to determine the required exposure time to impart a dose of 10 Gy. The x-ray output was investigated when the exposure failed to produce cell cycle arrest. The x-ray tube was fitted with only a 0.76 mm beryllium filter, and the spectrum therefore contained a high proportion of low energy photons which were being removed by the polystyrene sample flask as demonstrated by dose measurements in air and through the sample flask. Incorporation of a 0.5 mm aluminium filter removed most of these low energy photons but greatly reduced the dose rate to 3.8 mGy s-1. The manufacturer's quoted dose rates from a lightly filtered tube are misleading: the contribution of the very low energy component of the spectrum to the dose is not relevant since it will be removed by a sample flask. (note)

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/49/N39/pmb4_5_n01.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
49
Journal Issue
5
Journal Page Range
p. N39-N46
ISSN
0031-9155
CODEN