The use of dosimetric systems based on color changes in PMMA is standardized by ISO/ASTM 51276:2002(E). In a general way, the evaluation of doses in these materials is done correlating the optical absorbance of the material and the absorbed dose. Usually, the absorbance is measured using a spectrophotometer at a wavelength suitably chosen. In the present work, it was performed a comparative study between the techniques of spectrophotometry and optical densitometry in dyed PMMA dosimeters type Harwell RED 4034 and PMMA samples commercially available in Brazil. The main results of this study were the determination of the radiochromic response for different samples of PMMA and the determination of the calibration conditions using a 60Co beam. The studied materials are suitable for use as routine dosimeters, covering the range of the 500Gy to 100kGy. A direct comparison between the red dyed PMMA and the Harwell RED 4034 Perspex showed that the stabilization time and the fading of the radiochromic information in red dyed PMMA samples is higher than in RED 4034. However, both effects occur on period of weeks, which does not preclude its routine use. Furthermore, the increase in absorbance due to exposure to radiation, determined by the difference between specific absorbance of the irradiated sample and a non-irradiated one, is significantly higher in PMMA with red dyed that the RED 4034. In fact, the amplitude of absorbance peak of a sample irradiated with red dyed PMMA 2kGy and read at 602nm has the same order of magnitude from a peak of RED 4034 irradiated with 10 kGy and read at 640nm. A direct consequence of this fact is the real possibility of using simpler equipment (optical densitometer in this case) to measure the optical absorbance and evaluates the absorbed dose. (author)

Files

43048826.pdf

Files (297.1 kB)

Name Size Download all
md5:0e13082973eb4e3ac6cc2cde14f258ba
297.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
8 p.
Report number
INIS-BR--10933

Conference

Title
International nuclear atlantic conference. Nuclear energy: new jobs for a better life; 17. ENFIR: Meeting on nuclear reactor physics and thermal hydraulics; 10. ENAN: Meeting on nuclear applications; 2. ENIN: Meeting on nuclear industry
Acronym
INAC 2011
Dates
24-28 Oct 2011
Place
Belo Horizonte, MG (Brazil)

INIS

Country of Publication
Brazil
Country of Input or Organization
Brazil
INIS RN
43048826
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BRAZIL; DENSITOMETERS; DOSEMETERS; DOSIMETRY; PMMA; RADIATION DETECTION; RADIATION DOSES; RADIATION MONITORING; SPECTROPHOTOMETRY
Descriptors DEC
DETECTION; DEVELOPING COUNTRIES; DOSES; ESTERS; LATIN AMERICA; MEASURING INSTRUMENTS; MONITORING; ORGANIC COMPOUNDS; ORGANIC POLYMERS; PHOTOMETERS; POLYACRYLATES; POLYMERS; POLYVINYLS; SOUTH AMERICA