Published May 2016 | Version v1
Journal article

Optical imaging module for astigmatic detection system

  • 1. Department of Mechanical Engineering, National Taiwan University, Taipei 10617, Taiwan (China)
  • 2. Institute of Physics, Academia Sinica, Taipei 11529, Taiwan (China)
  • 3. Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig (Germany)

Description

In this paper, an optical imaging module design for an astigmatic detection system (ADS) is presented. The module is based on a commercial optical pickup unit (OPU) and it contains a coaxial illuminant for illuminating a specimen. Furthermore, the imaging module facilitates viewing the specimen and the detection laser spot of the ADS with a lateral resolution of approximately 1 μm without requiring the removal of an element of the OPU. Two polarizers and one infrared filter are used to eliminate stray laser light in the OPU and stray light produced by the illuminant. Imaging modules designed for digital versatile disks (DVDs) and Blu-ray DVDs were demonstrated. Furthermore, the module can be used for imaging a small cantilever with approximate dimensions of 2 μm (width) × 5 μm (length), and therefore, it has the potential to be used in high-speed atomic force microscopy.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
87
Journal Issue
5
Journal Page Range
vp.
ISSN
0034-6748
CODEN
RSINAK

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48042575
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
APPROXIMATIONS; ATOMIC FORCE MICROSCOPY; DESIGN; DETECTION; FILTERS; IMAGES; LASERS; LENGTH; RESOLUTION; VELOCITY
Descriptors DEC
CALCULATION METHODS; DIMENSIONS; MICROSCOPY

Optional Information

Notes
(c) 2016 Author(s)