Published April 1992 | Version v1
Journal article

Fabrication of tissue compensators for megavoltage beams using tissue equivalent materials

  • 1. Tata Memorial Hospital, Bombay (India). Dept. of Medical Physics

Description

Irregular body surface of the patient give rise to inhomogeneous dose distributions around the target volume. Homogeneous dose distribution can be obtained using tissue compensators. Preparation of tissue compensators with aluminium or lead blocks is a tedious method and commercially available devices are expensive. A compensator device has been designed to use a tissue equivalent material to be kept at shield tray level, which accounts for divergence of radiation beam. The reduced size of such tissue compensator obtained at much longer distance from the skin exhibits minimum risk of electron contamination. Experimental verifications were carried out for a tissue compensator by determining transmission at different off axis points for confirmation of the accuracy of the device. This paper presents the design of tissue compensator device and methods to prepare the tissue compensators using tissue equivalent materials. (author). 6 refs., 3 figs

Additional details

Publishing Information

Journal Title
AMPI (Association of Medical Physicists of India) Medical Physics Bulletin
Journal Volume
17
Journal Issue
2
Journal Page Range
p. 1-3.
ISSN
0250-5002
CODEN
AMPBDV