A new dynamic myocardial phantom for the assessment of left ventricular function by gated single-photon emission tomography
Creators
- 1. Dept. of Radiological Technology, Coll. of Medical Technology, Hokkaido Univ., Sapporo (Japan)
- 2. Dept. of Nuclear Medicine, School of Medicine, Hokkaido Univ., Sapporo (Japan)
- 3. Dept. of Tracer Kinetics, Hokkaido Univ., Sapporo (Japan)
Description
Gated myocardial perfusion single-photon emission tomography (SPET) has been used for the measurement of left ventricular (LV) function and validated by means of comparison with other imaging modalities. We have designed a new dynamic myocardial phantom in order to validate the LV function as assessed by the use of gated myocardial perfusion SPET. The phantom consists of two half-ellipsoids (an endocardial surface and an epicardial surface) and a thorax. The myocardial space is filled with a radioactive solution. The endocardial surface moves continuously towards and away from the epicardial surface in the longitudinal axis to vary the LV volume [143 ml at end-diastole (ED), 107 ml at end-systole (ES)] and thickness (apex 8 mm at ED and 26 mm at ES, midplane 8 mm). The mean values of wall motion (WM) for the apical midplane region and the basal midplane region were 5 mm and 2 mm, respectively. Gated myocardial SPET was performed during 8 and 16 intervals. These projection data sets were processed using a Butterworth filter with an order of 5 and a critical frequency of 0.34 cycles/cm. LV function was calculated using the quantitative gated SPET (QGS) algorithm. The LV function values estimated by gated SPET during 16 intervals [22% for ejection fraction (EF), 3.7 mm for WM of the apical midplane, 1.7 mm for WM of the basal midplane] closely resembled actual LV functions [25% for EF, 5 mm for WM of the apical midplane, 2 mm for WM of the basal midplane]. However, the estimated values during 8 intervals were smaller than those during 16 intervals (19% for EF, 3.3 mm for WM of the apical-midplane, 1.1 mm for WM of the basal-midplane). The estimated LV volumes closely correlated with the actual volumes (r=0.99 for 16 intervals, r=0.95 for 8 intervals). Utilizing this phantom, LV function estimated using gated myocardial SPET can be compared with actual values. (orig.)
Additional details
Publishing Information
- Journal Title
- European Journal of Nuclear Medicine
- Journal Volume
- 27
- Journal Issue
- 10
- Journal Page Range
- p. 1525-1530
- ISSN
- 0340-6997
- CODEN
- EJNMD9
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 31063293
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOLOGICAL FUNCTIONS; DYNAMIC FUNCTION STUDIES; MYOCARDIUM; PHANTOMS; RADIOPHARMACEUTICALS; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; TISSUE DISTRIBUTION; UPTAKE; VIABILITY
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISTRIBUTION; DRUGS; EMISSION COMPUTED TOMOGRAPHY; HEART; LABELLED COMPOUNDS; MATERIALS; MOCKUP; MUSCLES; ORGANS; RADIOACTIVE MATERIALS; STRUCTURAL MODELS; TOMOGRAPHY