Published November 21, 2009 | Version v1
Journal article

An alternative method for noise analysis using pixel variance as part of quality control procedures on digital mammography systems

  • 1. National Expert and Training Centre, Radboud University Nijmegen Medical Centre (LRCB), Weg door Jonkerbos 90, 6532 SZ Nijmegen (Netherlands)
  • 2. NCCPM, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom)
  • 3. Medical Physics Department, General Hospital of Pistoia, Viale Matteotti 19, 51100 Pistoia (Italy)
  • 4. Referenzzentrum Mammographie, Strasse des 17 Juni 106-108, D-10623 Berlin (Germany)

Description

According to the European Guidelines for quality assured breast cancer screening and diagnosis, noise analysis is one of the measurements that needs to be performed as part of quality control procedures on digital mammography systems. However, the method recommended in the European Guidelines does not discriminate sufficiently between systems with and without additional noise besides quantum noise. This paper attempts to give an alternative and relatively simple method for noise analysis which can divide noise into electronic noise, structured noise and quantum noise. Quantum noise needs to be the dominant noise source in clinical images for optimal performance of a digital mammography system, and therefore the amount of electronic and structured noise should be minimal. For several digital mammography systems, the noise was separated into components based on the measured pixel value, standard deviation (SD) of the image and the detector entrance dose. The results showed that differences between systems exist. Our findings confirm that the proposed method is able to discriminate systems based on their noise performance and is able to detect possible quality problems. Therefore, we suggest to replace the current method for noise analysis as described in the European Guidelines by the alternative method described in this paper.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/54/22/004

Additional details

Identifiers

DOI
10.1088/0031-9155/54/22/004;
PII
S0031-9155(09)24293-X;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
54
Journal Issue
22
Journal Page Range
p. 6809-6822
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41062309
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DIGITAL SYSTEMS; MAMMARY GLANDS; NEOPLASMS; NOISE; QUALITY CONTROL; RECOMMENDATIONS; SCREENING
Descriptors DEC
BODY; CONTROL; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY