An alternative method for noise analysis using pixel variance as part of quality control procedures on digital mammography systems
- 1. National Expert and Training Centre, Radboud University Nijmegen Medical Centre (LRCB), Weg door Jonkerbos 90, 6532 SZ Nijmegen (Netherlands)
- 2. NCCPM, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom)
- 3. Medical Physics Department, General Hospital of Pistoia, Viale Matteotti 19, 51100 Pistoia (Italy)
- 4. Referenzzentrum Mammographie, Strasse des 17 Juni 106-108, D-10623 Berlin (Germany)
Description
According to the European Guidelines for quality assured breast cancer screening and diagnosis, noise analysis is one of the measurements that needs to be performed as part of quality control procedures on digital mammography systems. However, the method recommended in the European Guidelines does not discriminate sufficiently between systems with and without additional noise besides quantum noise. This paper attempts to give an alternative and relatively simple method for noise analysis which can divide noise into electronic noise, structured noise and quantum noise. Quantum noise needs to be the dominant noise source in clinical images for optimal performance of a digital mammography system, and therefore the amount of electronic and structured noise should be minimal. For several digital mammography systems, the noise was separated into components based on the measured pixel value, standard deviation (SD) of the image and the detector entrance dose. The results showed that differences between systems exist. Our findings confirm that the proposed method is able to discriminate systems based on their noise performance and is able to detect possible quality problems. Therefore, we suggest to replace the current method for noise analysis as described in the European Guidelines by the alternative method described in this paper.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/22/004Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/22/004;
- PII
- S0031-9155(09)24293-X;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 22
- Journal Page Range
- p. 6809-6822
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41062309
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DIGITAL SYSTEMS; MAMMARY GLANDS; NEOPLASMS; NOISE; QUALITY CONTROL; RECOMMENDATIONS; SCREENING
- Descriptors DEC
- BODY; CONTROL; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY