Well-type HPGe-detector absolute-peak-efficiency calibration and true-coincidence correction
Creators
Description
A personal-computer-based program SWELL has been developed for well-type HPGe detector effective-solid-angle calculation. This program is very useful in constructing the absolute-peak efficiency (εp) versus γ-ray energy (Eγ) curves for different sample geometries based on a pre-determined εp under a reference counting geometry. The validity of using this program for εp(Eγ) conversion was successfully demonstrated for photons in the energy range ∼20 keV-1.5 MeV; the overall uncertainty can be controlled to be within 3%. In addition, a semi-empirical method has been developed to estimate the true-coincidence correction (COI) factor for well-type HPGe detector. Results based on 60Co, 139Ce, 133Ba and 59Fe sources indicated that the estimated COI factors are in good agreement with the experimentally validated COI values
Additional details
Identifiers
- PII
- S0168900298014284;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 425
- Journal Issue
- 3
- Journal Page Range
- p. 504-515
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Syrian Arab Republic
- INIS RN
- 34064027
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CORRECTIONS; EFFICIENCY; GAMMA RADIATION; HIGH-PURITY GE DETECTORS; KEV RANGE 10-100; MEV RANGE 01-10; PHOTONS; S CODES
- Descriptors DEC
- BOSONS; COMPUTER CODES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ENERGY RANGE; GE SEMICONDUCTOR DETECTORS; IONIZING RADIATIONS; KEV RANGE; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTORS; RADIATIONS; SEMICONDUCTOR DETECTORS
Optional Information
- Copyright
- Copyright (c) 1999 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.