Description

A home intercomparison of emissivity measurements with Ge(Li) detector and 152Eu source was organized by National Institute of Metrology (NIM). 152Eu standard source was used for the efficiency calibration of 130 cm3 Ge(Li) detector. The emissivity rate of gamma rays of 22Na (1274.5 KeV), 54Mn (834.85 KeV), 57Co (122.06 KeV), 60Co (1173.22 KeV, 1332 KeV), 134Cs (604.66 KeV, 795.76 KeV), 137Cs (661.65 KeV) and 133Ba (356.04 KeV) were measured with estimated errors less than 2%. Our results are compared with results of center of English Radiochemistry as well as those of NIM