Published April 1980
| Version v1
Journal article
Multi-anode resistive wire proportional counter as a focal plane detector
- 1. Tokyo Univ. (Japan). Inst. for Nuclear Study
- 2. Hiroshima Univ. (Japan). Dept. of Applied Physics and Chemistry
Description
A resistive-wire position-sensitive proportional counter used in the focal plane of a magnetic spectrograph has been improved by a multi-anode arrangement available for the non-normal incidence of particles. The achieved position resolution is 0.75 mm fwhm for 28 MeV protons and 0.9 mm fwhm for 50 MeV protons with an incidence angle of 350 to the wire axis. The non-normal incidence effect is also described with a measurement of the angle dependence of the position resolution in a simple resistive-wire proportional counter. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods
- Journal Volume
- 171
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 61-66
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 11555514
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CHARGED PARTICLE DETECTION; DIRECTIONAL RADIATION DETECTOR; MAGNETIC SPECTROMETERS; MEV RANGE 10-100; MULTIWIRE PROPORTIONAL CHAMBER; POSITION SENSITIVE DETECTORS; PROTON DETECTION; SPATIAL RESOLUTION
- Descriptors DEC
- ENERGY RANGE; MEASURING INSTRUMENTS; MEV RANGE; PROPORTIONAL COUNTERS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SPECTROMETERS