Performance evaluation of commercial radionuclide calibrators in Indonesian hospitals

  • 1. Center for Technology of Radiation Safety and Metrology, National Nuclear Energy Agency, P.O. Box 7043-JKSKL, Jakarta Selatan 12070 (Indonesia)

Description

Dose calibrator is one of the supporting equipments in the field of nuclear medicine. At the hospitals, dose calibrator is used for activity measurement of radiopharmaceutical before it is administered to patients. Comparison of activity measurements of 131I and 99mTc with dose calibrators was organized in Indonesia during 2007–2010 with the the aim of obtaining information dose calibrator performance in the hospitals. Seven Indonesian hospitals participated in this comparison. The measurement results were evaluated using the En criteria. The result presented in this paper facilitated the evaluation of dose calibrator performance at several hospitals. - Highlights: ► National comparisons of 131I and 99mTc radionuclides in Indonesian hospitals. ► Standardization using a Centronic IG11/A20 4πγ Ionization Chamber and participants using commercial radionuclide calibrators. ► Performance radionuclide calibrator in nuclear medicine in Indonesia. ► Measurement of activity of 99mTc and 131I was found satisfactory.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2012.02.113

Additional details

Identifiers

DOI
10.1016/j.apradiso.2012.02.113;
PII
S0969-8043(12)00161-3;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
70
Journal Issue
9
Journal Page Range
p. 2243-2245
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
18. international conference on radionuclide metrology and its applications
Dates
19-23 Sep 2011
Place
Tsukuba (Japan)

Optional Information

Copyright
Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.