Determination of the energy response of dosimeters with a Bragg-monochromatized beam

  • 1. Atominstitut der Oesterreichischen Universitaeten, Vienna

Description

In the energy region below 50 keV the energy response of dosimeters is usually determined by means of filtered bremsstrahlung radiation. The 'monochromatic' radiation produced by filtering the continuous X-ray spectrum of an X-ray tube shows a broad energy distribution for the maximum of the intensity and a considerable 'tailing' to low energies. To avoid this disadvantage we used a Bragg monochromator, a commercially bremsspectrum of a tungsten X-ray tube. The angle of reflection and the energy are determined by the Bragg condition. The energy resolution obtained was 2%. The photon flux density at the position of the dosimeter was measured with an efficiency calibrated Ge(Li) semiconductor detector. From this photon flux density the actual exposure of the dosimeter under examination can be calculated. (author)

Additional details

Publishing Information

Journal Title
Acta Phys. Acad. Sci. Hung.
Journal Volume
52
Journal Issue
3-4
Series
Acta Phys. Acad. Sci. Hung.
Journal Page Range
333-339
ISSN
0001-6705

Conference

Title
Hungarian-Austrian health physicists' meeting.
Dates
28 Sep - 1 Oct 1981.
Place
Gyoer (Hungary).

INIS

Country of Publication
Hungary
Country of Input or Organization
Hungary
INIS RN
14784067
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BRAGG REFLECTION; CALIBRATION; DOSEMETERS; DOSIMETRY; ENERGY DEPENDENCE; ENERGY RESOLUTION; KEV RANGE 10-100; MONOCHROMATORS; RADIATION DOSES; X-RAY DIFFRACTION
Descriptors DEC
COHERENT SCATTERING; DIFFRACTION; ENERGY RANGE; KEV RANGE; MEASURING INSTRUMENTS; REFLECTION; RESOLUTION; SCATTERING

Optional Information

Notes
4 refs.