Determination of the energy response of dosimeters with a Bragg-monochromatized beam
Description
In the energy region below 50 keV the energy response of dosimeters is usually determined by means of filtered bremsstrahlung radiation. The 'monochromatic' radiation produced by filtering the continuous X-ray spectrum of an X-ray tube shows a broad energy distribution for the maximum of the intensity and a considerable 'tailing' to low energies. To avoid this disadvantage we used a Bragg monochromator, a commercially bremsspectrum of a tungsten X-ray tube. The angle of reflection and the energy are determined by the Bragg condition. The energy resolution obtained was 2%. The photon flux density at the position of the dosimeter was measured with an efficiency calibrated Ge(Li) semiconductor detector. From this photon flux density the actual exposure of the dosimeter under examination can be calculated. (author)
Additional details
Publishing Information
- Journal Title
- Acta Phys. Acad. Sci. Hung.
- Journal Volume
- 52
- Journal Issue
- 3-4
- Series
- Acta Phys. Acad. Sci. Hung.
- Journal Page Range
- 333-339
- ISSN
- 0001-6705
Conference
- Title
- Hungarian-Austrian health physicists' meeting.
- Dates
- 28 Sep - 1 Oct 1981.
- Place
- Gyoer (Hungary).
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 14784067
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BRAGG REFLECTION; CALIBRATION; DOSEMETERS; DOSIMETRY; ENERGY DEPENDENCE; ENERGY RESOLUTION; KEV RANGE 10-100; MONOCHROMATORS; RADIATION DOSES; X-RAY DIFFRACTION
- Descriptors DEC
- COHERENT SCATTERING; DIFFRACTION; ENERGY RANGE; KEV RANGE; MEASURING INSTRUMENTS; REFLECTION; RESOLUTION; SCATTERING
Optional Information
- Notes
- 4 refs.