Published November 15, 1987 | Version v1
Journal article

Long term stability of collected charge in tetramethylsilane

  • 1. Saitama Coll. of Health, Urawa (Japan)
  • 2. Waseda Univ., Tokyo (Japan). Science and Engineering Research Lab.

Description

The long term stability of ionization yield in liquid tetramethylsilane (TMS) has been studied with a parallel plate ionization chamber. The collected charge due to conversion electrons from a 207Bi source was almost constant over one year if TMS was purified several times and the chamber was baked out well. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
261
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
598-599
ISSN
0168-9002
CODEN
NIMAE

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
19029010
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ELECTRIC CHARGES; IONIZATION; IONIZATION CHAMBERS; LIQUIDS; PERFORMANCE; SILANES; STABILITY
Descriptors DEC
FLUIDS; HYDRIDES; HYDROGEN COMPOUNDS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SILICON COMPOUNDS