Published September 1982 | Version v1
Journal article

Optimization of ionizing radiation usage in medical imaging by means of image enhancement techniques

Creators

  • 1. Capintec, Inc., Ramsey, NJ

Description

This paper focuses attention on the fact that images considered unsatisfactory for medical analysis may be rendered usable following various image enhancement techniques. In those cases where this form of image processing can sufficiently improve otherwise unusable images, the necessity of repetitive administration of radionuclides to patients may be avoided with a concomitant reduction in ionizing radiation dosages. Application of these techniques in a clinical situation reduced the dose in thallium studies by 50%. Similar results may be obtainable with other nuclear medicine procedures

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
9
Journal Issue
5
Series
Med. Phys.
Journal Page Range
733-737
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
16077823
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
HEART; IMAGE INTENSIFIERS; IMAGE PROCESSING; PATIENTS; RADIATION DOSES; SCINTISCANNING; SKELETON; THALLIUM ISOTOPES
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; COUNTING TECHNIQUES; DIAGNOSTIC TECHNIQUES; ORGANS; RADIOISOTOPE SCANNING