Published May 8, 2007 | Version v1
Journal article

On-line arsenic co-precipitation on ethyl vinyl acetate turning-packed mini-column followed by hydride generation-ICP OES determination

  • 1. Consejo Nacional de Investigaciones Cientificas y Tecnicas (CONICET), Rivadavia 1917, CP C1033 AAJ, Ciudad de Buenos Aires (Argentina)
  • 2. Department of Analytical Chemistry, Faculty of Chemistry, Biochemistry and Pharmacy, National University of San Luis, Chacabuco and Pedernera, P.O. Box 375, 5700 San Luis (Argentina)
  • 3. Department of Analytical Chemistry, Faculty of Chemistry, Biochemistry and Pharmacy, National University of San Luis, Chacabuco and Pedernera, P.O. Box 375, 5700 San Luis (Argentina) and Consejo Nacional de Investigaciones Cientificas y Tecnicas (CONICET), Rivadavia 1917, CP C1033 AAJ, Ciudad de Buenos Aires (Argentina)

Description

An alternative and new system for on-line preconcentration of arsenic by sorption on a mini-column associated to hydride generation - inductively coupled plasma - optical emission spectrometry determination was studied. It is based on the sorption of arsenic on a column packed with ethyl vinyl acetate (EVA) turnings and the use of La(III) as co-precipitant reagent. This polymeric material was employed here for the first time as filling material for column preconcentration. It could work both as adsorbent and as sieve material. Sample and co-precipitant agent (lanthanum nitrate) were off-line mixed and merged with ammonium buffer solution (pH 10.0), which promoted precipitation and quantitative collection on the small EVA turnings. The arsenic preconcentrated by co-precipitation with lanthanum hydroxide precipitate was subsequently eluted with hydrochloric acid, which was the medium used for hydride generation. Considering a flow rate of 5 ml/min, three enrichment factors were obtained, 28-, 38- and 45-fold at three different sampling times, 60, 120 and 180 s; respectively. The detection limits (3 s) obtained for each case were 0.013, 0.009 and 0.007 μg/l. Additionally, the calculated precisions expressed as relatively standard deviation (R.S.D.) were 0.9, 1.3 and 1.1%. Satisfactory results were obtained for the determination of arsenic in standard reference material NIST 1643e Trace Elements in Water and drinking water samples

Additional details

Identifiers

DOI
10.1016/j.jhazmat.2006.09.051;
PII
S0304-3894(06)01120-4;

Publishing Information

Journal Title
Journal of Hazardous Materials
Journal Volume
143
Journal Issue
1-2
Journal Page Range
p. 431-436
ISSN
0304-3894
CODEN
JHMAD9

Optional Information

Copyright