Effects of dislocations and crystal defects on the energy resolution and response uniformity of mercuric iodide detectors
Description
The energy resolution and response uniformity of HgI2 detectors to nuclear radiation depends strongly on dislocation density and distribution in the HgI2 crystal bulk. Several large planar detectors fabricated from the interiors of large single crystals verified this dependence. Scanning with collimated low-energy (6 keV) and medium-energy (60 keV) gamma-rays confirmed the association of poor resolution and counting efficiency with regions of high dislocation density and crystal imperfection
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 12 p.
- Report number
- EGG--1183-2344
Conference
- Title
- Nuclear science, scintillation and semiconductor counter symposium.
- Dates
- 20 Oct 1976.
- Place
- New Orleans, Louisiana, United States of America (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 8318938
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CRYSTAL DEFECTS; DISLOCATIONS; EFFICIENCY; ENERGY RESOLUTION; GAMMA DETECTION; HGI2 SEMICONDUCTOR DETECTORS; PERFORMANCE; SENSITIVITY
- Descriptors DEC
- CRYSTAL STRUCTURE; LINE DEFECTS; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS
Optional Information
- Notes
- Available from NTIS. $3.50.
- Secondary number(s)
- CONF-761006--29.