Published January 2014 | Version v1
Journal article

Performance of flowing sample neutron activation analysis technique for determination of multi-elemental content

  • 1. Atomic Energy Authority, Cairo (Egypt). Egypt Second Research Reactor
  • 2. Atomic Energy Authority, Cairo (Egypt). Laboratories for Detection of Landmines and Illicit Materials, Nuclear Research Centre

Description

In order to measure and obtain reliable data about elements via measuring of short-lived isotopes in liquid samples using neutron activation analysis, experiments have been carried out using a flowing sample neutron activation analysis method. During this work the performance and reliability for the proposed method have been tested using synthetic multi elements liquid samples. The radionuclide, associated elements and their concentration level have been measured using a counting system with a fixed dead-time. An HPGe spectrometry system was used to determine the most important γ-ray energies and intensities of these isotopes. The obtained results demonstrated the reliability and accuracy (bias <5 % and zeta score <2) of the tested method for determination the elemental content of liquid samples. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
299
Journal Issue
1
Journal Page Range
p. 89-93
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
10 refs.