Published July 2009 | Version v1
Journal article

Validation of the determination of tin by k0-instrumental neutron activation analysis in foodstuff

  • 1. Belgian Nuclear Research Centre, SCK EN, Mol (Belgium)
  • 2. Nuclear Technology Development Center/Brazilian Commission for Nuclear Energy (CDTN/CNEN), Belo Horizonte, MG (Brazil). Laboratory for Neutron Activation Analysis

Description

Several recent intercomparisons and certification exercises have shown that the determination of tin in e.g. food samples and plastics is not straightforward. k0-Instrumental Neutron Activation Analysis (k0-INAA) offers some intrinsic quality control opportunities for the determination of the tin content in samples since several monitoring radioisotopes are formed. In this work we will discuss the validation of the determination of tin by k0-INAA using different reference materials. The results show that only few of these radioisotopes are reliable in terms of accuracy and that from a survey of the tin content in a range of canned foods detection limits as low as 2 mg/kg can be achieved. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
281
Journal Issue
1
Journal Page Range
p. 35-39
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
16 refs.