Background
Clinically applied radiopharmaceuticals have to meet quality release criteria like a high radiochemical yield and radiochemical purity. Many radiopharmaceuticals do not have marketing authorization and have no dedicated monograph within the European pharmacopeia, therefore general monographs on quality control have to be applied for clinical applications. These criteria require standardization and validation in labeling and preparation, including QC measurements according to well-defined standard operation procedures. QC measurements however, are often based on detection techniques specific for a certain LC-system. Multi-institutional research and development of new radiopharmaceuticals lead to an increase in multicenter trials. Although all institutes' radiopharmacies are using the same standardized labeling and operation procedures, they often use different LC and radiodetection systems. Here we present a comparison of QC assessments for 3 radiopharmaceuticals with focus on the interpretation of chromatograms, data-output and potential differences in local practical performances of QC on (U)HPLC.Methods
QC assessments for [111In]In-CCK, [68Ga]Ga-Bombesin and [177Lu]Lu-PSMA analogs were compared. Two of the radiopharmaceutical QC assessments were also applied in other institutes using their own HPLC-systems and concordant software. Data from the HPLC-injections and measurements is processed and summarized in chromatograms, based on a variety of smoothing algorithms for which different software programs are applied. Described radiopeptides were labeled and analyzed according their standardized labeling and operation procedures.
Results
Integration of main peaks on chromatograms resulted in a range of RCP, depending on the smoothing algorithm used. [111In]In-CCK(A), 68Ga-Bombesin(B) and [177Lu]Lu-PSMA(C) analogs had a RCP range of 88%–96%(A), 89–95%(B) and 92–99%(C) respectively. Important factors affecting final RCP value were site specific background radiation-levels, intrinsic system properties such as noise and sensitivity, personal interpretation e.g. peak-tailing and smoothing algorithms.
Conclusion
Measurement of RCP shows a strong method- and system-dependency, even when parameters are validated, standardized and SOP are followed. Release criteria are frequently based on RCP data from one central location. The lack of inter inter institutional validation and standardization in RCP determination makes the results therefore rather arbitrary. For multicenter trials, we recommend to compare locally determined RCP under validated and standardized conditions of in-line activity detection between institutes for each radiopharmaceutical.
Additional details
Identifiers
Publishing Information
- Journal Title
- EJNMMI Radiopharmacy and Chemistry
- Journal Volume
- 4
- Journal Issue
- 1
- Journal Page Range
- p. 1-13
- ISSN
- 2365-421X
INIS
- Country of Publication
- Switzerland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54096568
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BACKGROUND RADIATION; COMPUTER CODES; GALLIUM 68; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; INDIUM 111; LUTETIUM 177; QUALITY CONTROL; RADIOCHEMISTRY; RADIOPHARMACEUTICALS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; CHEMISTRY; CHROMATOGRAPHY; CONTROL; DAYS LIVING RADIOISOTOPES; DRUGS; ELECTRON CAPTURE RADIOISOTOPES; GALLIUM ISOTOPES; HOURS LIVING RADIOISOTOPES; INDIUM ISOTOPES; INTERMEDIATE MASS NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LABELLED COMPOUNDS; LIQUID COLUMN CHROMATOGRAPHY; LUTETIUM ISOTOPES; MATERIALS; MATHEMATICAL LOGIC; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATIONS; RADIOACTIVE MATERIALS; RADIOISOTOPES; RARE EARTH NUCLEI; SEPARATION PROCESSES
Optional Information
- Copyright
- Copyright (c) 2019 The Author(s)