Gamma-ray imaging performance of the GRI+ Compton camera
Creators
- 1. Oliver Lodge Laboratory, University of Liverpool, Liverpool, L69 7ZE (United Kingdom)
- 2. STFC Daresbury Laboratory, Daresbury, Warrington, WA4 4AD (United Kingdom)
Description
GRI+ (Gamma-Ray Imager) is a cart-based Compton camera developed at the University of Liverpool. The system comprises three electrically-cooled semiconductor detectors; two planar double-sided strip detectors (Si(Li) and HPGe) and an HPGe coaxial detector. The mobile cart houses the necessary components for data acquisition and analysis, providing complete functionality for imaging gamma-ray sources in an industrial setting. This paper reports on the operation of the system and the methods used to improve intrinsic system performance. The processing chain for extracting information pertaining to gamma-ray interactions has been developed and the imaging performance is characterised using an analytical reconstruction code. The obtainable Angular Resolution Measure (ARM) from a Cs-137 point source placed 10.5 cm from the front of the Si(Li) cryostat was determined to be 4.12∘ after the application of data processing techniques. The system energy resolution, which is key to isotope identification and image quality, was calculated to be less than 0.5% at 1332.5 keV. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/16/09/P09015Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 16
- Journal Issue
- 09
- Journal Page Range
- [15 p.]
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53083538
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CAMERAS; CESIUM 137; CRYOSTATS; DATA ACQUISITION; ENERGY RESOLUTION; GAMMA RADIATION; HIGH-PURITY GE DETECTORS; KEV RANGE; PERFORMANCE; POINT SOURCES
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; CONTROL EQUIPMENT; DATA PROCESSING; ELECTROMAGNETIC RADIATION; ENERGY RANGE; EQUIPMENT; GE SEMICONDUCTOR DETECTORS; INTERMEDIATE MASS NUCLEI; IONIZING RADIATIONS; ISOTOPES; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; PROCESSING; RADIATION DETECTORS; RADIATION SOURCES; RADIATIONS; RADIOISOTOPES; RESOLUTION; SEMICONDUCTOR DETECTORS; THERMOSTATS; YEARS LIVING RADIOISOTOPES