Comparison of alternative methods for multiplet deconvolution in the analysis of gamma-ray spectra

Description

Three methods for multiplet deconvolution were tested using the 1995 IAEA reference spectra: Total area determination, iterative fitting and the library-oriented approach. It is concluded that, if statistical control (i.e. the ability to report results that agree with the known, true values to within the reported uncertainties) is required, the total area determination method performs the best. If high deconvolution power is required and a good, internally consistent library is available, the library oriented method yields the best results. Neither Erdtmann and Soyka's gamma-ray catalogue nor Browne and Firestone's Table of Radioactive Isotopes were found to be internally consistent enough in this respect. In the absence of a good library, iterative fitting with restricted peak width variation performs the best. The ultimate approach as yet to be implemented might be library-oriented fitting with allowed peak position variation according to the peak energy uncertainty specified in the library. (author)