Published May 1981 | Version v1
Journal article

Accuracy of some routine method used in clinical chemistry as judged by isotope dilution-mass spectrometry

Description

Serum from patients was pooled, filtered, dispensed, and frozen. This pooled specimen was used for accuracy control in 64 participating laboratories in Sweden. Mean values (state-of-the-art values) were obtained for creatinine, cholesterol, glucose, urea, uric acid, and cortisol. These values were compared with values obtained with highly accurate reference methods based on isotope dilution-mass spectrometry. Differences were marked in the case of determination of creatinine and cortisol. Concerning the other components, the differences between the state-of-the-art value and the values obtained with the reference methods were negligible. Moreover, the glucose oxidase and the oxime methods for determination of glucose and urea were found to give significantly lower values than the hexokinase and urease methods, respectively. Researchers conclude that methods with a higher degree of accuracy are required for routine determination of creatinine and cortisol

Additional details

Publishing Information

Journal Title
Clin. Chem. (Winston-Salem, N.C.)
Journal Volume
27
Journal Issue
5
Series
Clin. Chem. (Winston-Salem, N.C.).
Journal Page Range
733-735
ISSN
0009-9147