Published November 2016 | Version v1
Journal article

Improved density profile measurements in the C-2U advanced beam-driven Field-Reversed Configuration (FRC) plasmas

  • 1. Tri Alpha Energy, Inc., P.O. Box 7010, Rancho Santa Margarita, California 92688 (United States)

Description

In the prior C-2 experiment, electron density was measured using a two-color 6-chord CO2/HeNe interferometer. Analysis shows that high-frequency common mode phase noise can be reduced by a factor of 3 by constructing a reference chord. In the system upgrade from C-2 to C-2U a 4-chord far-infrared laser interferometer was developed, which demonstrated superior sensitivity (1 × 1016 m−2 at >1 MHz bandwidth) and solved the under spatial sampling issue of the C-2 interferometer system. Improved density-profile measurement results are presented in this paper, including evidence of fast-ion modified density profile and stabilization of the n = 1 plasma wobble mode.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
87
Journal Issue
11
Journal Page Range
vp.
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2016 Author(s)