Published September 2011 | Version v1
Journal article

Direct electrochemistry of myoglobin in a layer-by-layer film on an ionic liquid modified electrode containing CeO2 nanoparticles and hyaluronic acid

  • 1. Institute of Analytical Science, Shaanxi Provincial Key Laboratory of Electroanalytical Chemistry, Northwest University, Xian, Shaanxi 710069 (China)

Description

We describe an ionic liquid modified electrode (CPE-IL) for sensing hydrogen peroxide (HP) that was modified by the layer-by-layer technique with myoglobin (Mb). In addition, the surface of the electrode was modified with CeO2 nanoparticles (nano-CeO2) and hyaluronic acid. UV-vis and FTIR spectroscopy confirmed that Mb retains its native structure in the composite film. Scanning electron microscopy showed that the nano-CeO2 closely interact with Mb to form an inhomogeneously distributed film. Cyclic voltammetry reveals a pair of quasi-reversible redox peaks of Mb, with the cathodic peak at -0. 357 V and the anodic peak at -0. 269 V. The peak separation (ΔEp) and the formal potential (Eσ) are 88 mV and -0. 313 V (vs. Ag/AgCl), respectively. The Mb immobilized in the modified electrode displays an excellent electrocatalytic activity towards HP in the 0. 6 to 78. 0 μM concentration range. The limit of detection is 50 nM (S/N = 3), and then the Michaelis-Menten constant is 71. 8 μM. We believe that such a composite film has potential to further investigate other redox proteins and in the fabrication of third-generation biosensors. (author)

Additional details

Identifiers

Publishing Information

Journal Title
Microchimica Acta (Online)
Journal Volume
174
Journal Issue
3-4
Journal Page Range
p. 273-280
ISSN
1436-5073