SU-F-T-573: Evaluation of EBT-XD Radiochromic Films for Verification of SBRT and SRS Treatment Delivery
Creators
- 1. Centre hospitalier de l'Universite de Montreal, Montreal, Quebec (Canada)
Description
Purpose: To evaluate the accuracy and precision of radiochromic films EBT-XD for quality control of stereotaxic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) plan delivery. Methods: A film cut in 3×10 cm2 strips was irradiated from 0 to 20 Gy in increments of 1 to 1.5 Gy using a 15 MeV electron beam. Triple-channel film calibration was completed 24 hours later by scanning the film strips on an Epson 10000XL scanner using a well-defined protocol. Several dose measurements of increasing complexity were subsequently performed with Varian iX accelerators. Pieces of films were first irradiated in a solid water phantom with 6 MV photon beams and a static gantry to doses spanning the calibration range, either in a single field or multiple fields setup. High dose (>15 Gy per fraction) IMRT plans were then measured. Finally films were irradiated with volumetric modulated arc therapy (VMAT) plans of lung and spinal lesions with prescribed doses per fraction between 8 and 20 Gy. The dose measured with the films was compared to the calculated dose from the Eclipse planning system using the Anisotropic Analytical Algorithm (AAA). Results: 77 dose measurements were compared to either ion chamber measurements or dose calculations (reference). The average dose difference between film measurements and reference was 0.7 % and the standard deviation was 1.3%. The maximum and minimum dose differences were +3.5% and −2% in the 4 Gy to 20 Gy range. Measured dose profiles of lung and vertebra treatment plans agreed very well with the calculations. Conclusion: EBT-XD films are a useful dosimeter for quality control of SBRT and SRS plan delivery. The measurement of a full 2D dose plane with high spatial resolution and acceptable dose accuracy make it an advantageous choice compared to other detectors such as ion chambers or diodes.
Additional details
Identifiers
- DOI
- 10.1118/1.4956758;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 43
- Journal Issue
- 6
- Journal Page Range
- vp.
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48097806
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ACCURACY; ARSENIC IONS; DELIVERY; DOSES; ELECTRON BEAMS; GY RANGE 01-10; GY RANGE 10-100; IONIZATION CHAMBERS; IRRADIATION; MEV RANGE 10-100; PHOTON BEAMS; PLANNING; QUALITY ASSURANCE; QUALITY CONTROL; RADIOTHERAPY; SPATIAL RESOLUTION
- Descriptors DEC
- ABSORBED DOSE RANGE; BEAMS; CHARGED PARTICLES; CONTROL; ENERGY RANGE; GY RANGE; IONS; LEPTON BEAMS; MEASURING INSTRUMENTS; MEDICINE; MEV RANGE; NUCLEAR MEDICINE; PARTICLE BEAMS; RADIATION DETECTORS; RADIATION DOSE RANGES; RADIOLOGY; RESOLUTION; THERAPY
Optional Information
- Notes
- (c) 2016 American Association of Physicists in Medicine