Efficacy of L-shaped shielding in interventional radiology by transradial approach and consideration of methods for appropriate use
- 1. Hamamatsu University Hospital, Department of Radiology, Hamamatsu, Shizuoka (Japan)
- 2. Takaoka City Hospital, Department of Radiology, Takaoka, Toyama (Japan)
- 3. Fujita Health University, School of Medical Sciences, Toyoake, Aichi (Japan)
Description
The present study investigated how effective an L-shaped shield was, depending on its position, in reducing a doctor's exposure to radiation during catheterization to access the transradial approach (TRA). The shield's effectiveness was evaluated by measuring the air kerma where the doctor stood under four conditions: with and without the shield, and with and without the shield in conjunction with conventional protection. To enable the shield to be positioned correctly in clinical practice, an illustrated instruction decal affixable to the shield's doctor-facing surface was produced, and the effectiveness of the decal was verified by means of a crossover test in which, as subjects of the study, different nurses set up the shield with and without the decal affixed to it. In the test, in which a human body phantom was used, the C-arm set at the PA angle, and the shield positioned 10 cm from the axilla of the phantom, the shield's effectiveness at 100 cm, 130 cm, and 160 cm above the floor where the doctor stood was 55%, 77%, and 47%, respectively. The effectiveness increased when the shield was positioned closer to the axilla. A significant difference in the positioning of the shield by the subjects was observed depending on whether or not the decal was affixed (p < 0.05, Wilcoxon signed-rank test), indicating that the use of the decal improved the positioning. It was concluded that, positioned correctly, the shield could effectively reduce the doctor's exposure to radiation during TRA. (author)
Availability note (English)
Available from DOI: https://doi.org/10.6009/jjrt.2023-1325Abstract (Japanese)
橈骨動脈アプローチを用いたIVRにおけるL型遮蔽板の効果と適切な使用方法を検討する.L型遮蔽板の有無,および従来の防護方法とL型遮蔽板を併用した場合での術者位置における空気カーマを測定することで遮蔽効果を評価した.適切な配置についてはL型遮蔽板の設置位置と遮蔽効果の関係から検討した.また臨床で適切に使用するために,L型遮蔽板に使用方法を記載したシートを貼付することを考案し,クロスオーバー試験を用いてその効果を検証した.Cアーム角度PAのとき,術者位置の床上100, 130, 160 cmの位置においてそれぞれ55%,77%,47%の遮蔽効果を示した.また設置する位置をファントム腋窩に近接させることで遮蔽効果は上昇した.考案したシートの有無で配置に有意な差を認め(p<0.05, Wilcoxon's signed rank test),シートを使用することで配置の改善が認められた.TRA用L型遮蔽板は適切に使用することで,効果的に術者の被ばくを低減することが可能である.(著者)Additional details
Additional titles
- Original title (Japanese)
- 橈骨動脈アプローチを用いたIVRにおけるL型遮蔽板の有効性と適正使用のための手法の検討
Identifiers
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi (Online)
- Journal Volume
- 79
- Journal Issue
- 3
- Series
- 雑誌名:日本放射線技術学会雑誌
- Journal Page Range
- p. 241-251
- ISSN
- 1881-4883
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 54097336
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CARDIOVASCULAR DISEASES; COMPARATIVE EVALUATIONS; CORONARIES; DOSEMETERS; FLUOROSCOPY; PHANTOMS; RADIATION DOSES; RADIATION PROTECTION; SCATTERING; SHIELDS
- Descriptors DEC
- ARTERIES; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; EVALUATION; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS
Optional Information
- Notes
- 26 refs., 13 figs., 3 tabs.